Logo

RPG Radiometer physics GmbH

ZC-Series Millimeter Wave Converters

Key Features:


 

Test Port ZC78 ZC90  ZC90E ZC110 ZC118 ZC140 ZC170 ZC220 ZC400 ZC500 ZC750 
Frequency Range [GHz] 53 to 78 60 to 90 60 to 90 75 to 110 78 to 118 90 to 140 110 to 170 140 to 220 170 to 260 220 to 330 260 to 400 325 to 500 500 to 750
Port Type WR3.6 x1.8 (UG387/U flange compatible) WR12 (UG387/U flange compatible) WR12(UG387/U flange compatible)

WR-2540

(UG387/U flange compatible)

WR2.4x1.2 (UG387/U flange compatible) WR-2032 (UG387/U flange compatible)

WM-1651 (UG387/U flange compatible)

WM-1295 (UG387/U flange compatible) WM-1092 (UG387/U flange compatible) WM-864 (UG387/U flange compatible) WM-710 (UG387/U flange compatible) WM-570 (UG387/U flange compatible) WM-380 (UG-387 flange compatible)
Output Power [dBm typ.] >+6,  +12 dBm (typ.)    

>+12 (n. trc.), typ. +14 dBm

>+6,  +9 dBm (typ.)

90 to 95 GHz >+5, +7 dBm (typ.),

   95 to 135 GHz >+7, +9 dBm (typ.), 

 135 to 140 GHz >+5, +7 dBm (typ.)

>+5 (n. trc.),  +9 dBm
140 to 145 GHz > -4 dBm (n.trc.), typ. 0 dBm
145 to 220 GHz > -2 dBm (n.trc.), typ. +2 dBm
170 to 240 GHz > -8 dBm, typ. -4 dBm
240 to 260 GHz >-12 dBm, typ. -7 dBm
220 to 320 GHz > -11 dBm (n.trc.), typ. -8 dBm
320 to 330 GHz > -12 dBm (n.trc.), typ. -9 dBm
> -15 dBm, typ. -10dBm

325 - 440 GHz > -17dBm, typ. -13 dBm

440 - 500 GHz

> -24 dBm, typ. -18 dBm

 
Output Power Attenuation [dB] 0 to 40 0 to 40 0 to 40 0 to 40 n.A. 0 to 40

0 to 40

0 to 40 0 to 40 0 to 40 0 to 40 0 to 40 0 to 40
Input Power Damage Level [dBm] +20 +20 +20 +20 +20 +20 +20 +20 +20 +20 +10 +10  
Stability (Magnitude [dB] / Phase [°] (typ.) ) typ. < 0.2 dB and typ. < 2°
typ. < 0.2 dB and typ. < 2°
typ. < 0.2 dB and typ. < 2°
typ. < 0.2 dB and typ. < 2° typ. < 0.2 dB and typ. < 2° typ. < 0.2dB and typ. < 2°

typ. < 0.3 dB and typ. < 4°

typ. < 0.3 dB and typ. < 4°

typ. < 0.4 dB and typ. < 4°

typ. < 0.4 dB and typ. < 6° typ. < 0.5 dB and typ. < 6° typ. < 0.5 dB and typ. < 6°  
Source Input (RF IN)      
Frequency Range [GHz] 13.25 to 19.5     12.5 to 18.33 13 to 19.66 15 to 23.33 9.16 to 14.16 11.66 to 18.33 14.16 to 21.66 12.22 to 18.33 14.44 to 22.22 9.02 to 13.88  
Port Type 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female
Input Power Range [dBm] -15 to +10 -15 to +10 -15 to +10 -15 to +10 -15 to +10 -15 to +10 -15 to +10 -15 to +10 -15 to +10 -15 to +10 -15 to +10 +5 to +10  
Local Oscillator Input (LO IN)      
Frequency Range [GHz] 8.83 to 13.00     9.37 to 13.75 13.00 to 19.66 11.21 to 17.46 10.97 to 16.97 11.66 to 18.31 14.16 to 21.66 9.15 to 13.75 13.00 to 20.00 13.53 to 20.982  
Port Type 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female 2.92 mm, female
Input Power Range [dBm] +5 to +10 +5 to +10 +5 to +10 +5 to +10 +5 to +10 +5 to +10 +5 to +10 +5 to +10 +5 to +10 +5 to +10 +5 to +10 +5 to +10 +5 to +10
Measurement Output (MEAS OUT)      
Frequency Range [MHz] 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000  
Port Type SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female  
Reference Output (REF OUT)      
Frequency Range [MHz] 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000 5 to 2000  
Port Type SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female SMA, female  
System Characteristics      

Source match 

(without system error correction)

> 19 dB,

typ. > 30 dB

    > 17 dB (n.trc.)1

> 19 dB,

typ. > 30 dB

> 25 dB (n.trc.)1 > 25 dB (n.trc.)1 > 25 dB (n.trc.)1 > 20 dB (n.trc.)1 > 20 dB (n.trc.)1 > 20 dB (n.trc.)1 > 20 dB (n.trc.)1 > 20 dB (n.trc.)1

Directivity

(without system error correction)

> 23 dB,

typ. > 37 dB

    > 21 dB (n.trc.)1

> 23 dB,

typ. > 30 dB

> 25 dB (n.trc.)1 > 25 dB (n.trc.)1 > 25 dB (n.trc.)1 > 20 dB (n.trc.)1 > 20 dB (n.trc.)1 > 20 dB (n.trc.)1 > 20 dB (n.trc.)1 > 20 dB (n.trc.)1
Dynamic Range > 100 dB, typ. 115 dB     > 110 dB , typ. 120 dB > 100 dB, typ. 110 dB > 105 dB, typ. 120 > 90 dB, typ. 105 dB > 100 dB, typ. 115 > 100 dB, typ. 110 > 100 dB, typ. 115 > 80 dB, typ. 95 > 85 dB, typ. 100  

 

    Dynamic range is defined as the difference between the data trace of the transmission magnitude with maximum test port output power and both
    test ports through-connected on the one hand and the RMS value of the data trace of the transmission magnitude produced by noise and crosstalk  
    with test ports short-circuited on the other. The specification is valid without system error correction and at 10Hz measurement bandwidth. The
    dynamic range can be increased by using a measurement bandwidth of 1Hz.
 
   1 Without consideration of measurement uncertainty.